Cookbook – Chem_Balance 1
Chemistry Lab Cookbook
Balance Software, v210.0
Instrument 1:
Cahn Balance
ODP #: 02344
Model #: 31
Website:
http://www.thermocahn.com
Software:
ODP Labview v.198
Instrument 2:
Scientech Balance
ODP #:
Model #: 202-004
Software:
ODP Labview v.198
Documentation Available:
Document Name/Title/Number
Source
Version
Associated
Instrument(s)
Associated
Measurement(s)
Cahn 28/29 Automatic Electrobalance W/
RS232 Output (Owner's Manual)
Cahn
Instruments,
Inc.
n/a
Cahn balances
Cahn Instruction Manual
Cahn
Instruments,
Inc.
Rev. 6/90
Cahn balances
Labview User Guide
In house
198
Zobrazit stránku 0
1 2 3 4 5 6 ... 14 15

Shrnutí obsahu

Strany 1 - Chemistry Lab Cookbook

Cookbook – Chem_Balance 1Chemistry Lab CookbookBalance Software, v210.0Instrument 1:Cahn BalanceODP #: 02344Model #: 31Website:http://www.thermoca

Strany 2 - Balance Systems User Guide

Cookbook – Chem_Balance 10detected by the beam position sensing system. A greater force on the beam will require a greater opposite forcefrom the torq

Strany 3 - Cookbook – Chem_Balance 3

Cookbook – Chem_Balance 11When the Weight Display has stabilized, push the CAL button once;Remove the calibration weight. You may now tare and make no

Strany 4

Cookbook – Chem_Balance 12To calibrate the Cahn balance using the software, first set the Stop Criteria to “By Statistics”; the ConfidenceInterval to

Strany 5 - Cookbook – Chem_Balance 5

Cookbook – Chem_Balance 13Select New to open the Balance_Calibration.vi window:Follow the on-screen instructions: Start with no weight on the sample p

Strany 6

Cookbook – Chem_Balance 14If the samples IDs have been downloaded directly from the database, scanned, or manually entered in either theCoulometer or

Strany 7 - Cookbook – Chem_Balance 7

Cookbook – Chem_Balance 15Document Information:Author, Original: Bill MillsVersion Number, Original: 179Revision History:210:Chris BennightSignature

Strany 8

Cookbook – Chem_Balance 2 Balance Systems User Guidev.210Accelerations, induced by the ship's motion, make precise weighing at sea difficult. To

Strany 9 - Cookbook – Chem_Balance 9

Cookbook – Chem_Balance 3A. 1. Principle and Mode of OperationThe Scientech balance system is composed of two Scientech top loading balances (model 2

Strany 10 - Cookbook – Chem_Balance 10

Cookbook – Chem_Balance 4A. 2. Tuning the SensorsOnce the Scientech.vi is loaded, the following Control window is displayed:Clicking on the Show Cali

Strany 11

Cookbook – Chem_Balance 5The tuner.vi allows the user to apply a gain and zero value to the reference sensor’s output so as to match the zeroand gain

Strany 12 - Select_Calibration.vi window:

Cookbook – Chem_Balance 6To calibrate the sensors, proceed as follows: Set the Test Method to “By Stats”; Set the Confidence Interval to 99.5% and Pre

Strany 13 - Cookbook – Chem_Balance 13

Cookbook – Chem_Balance 7Clicking on the Halt button will stop the process at any time;When the Test Method is switched to "By Counts", weig

Strany 14

Cookbook – Chem_Balance 8Select the Confidence Interval and Precision;Place a piece of weighing paper or a weighing boat on each balance and click on

Strany 15 - Document Information:

Cookbook – Chem_Balance 9B. Cahn MicrobalanceThe Cahn microbalance (model C-30/31) is designed for weights up to 3.5g and is sensitive to changes as s

Komentáře k této Příručce